944i Dielectric Tester

If you need one unit that does the work of a multitude of test instruments, consider the speed, power and flexibility of the Vitrek 944i.

The Vitrek 944i combines AC dielectric strength testing (AC Hipot) with DC dielectric strength testing (DC Hipot), insulation resistance (to 65 Gigohms) and earth continuity measurement all in one feature-rich automated instrument. This compact dielectric analyzer lets you measure four electrical safety test parameters from a single test connection-all with the touch of a button.

The Vitrek 944i delivers top performance and versatility for manufacturing test, product safety engineering, component evaluation, as well as research and development applications. You'll find that the multi-function 944i is loaded with features that even the larger more expensive dedicated testers don't offer. Powerful features including automatic calibration, built-in diagnostics, internal "watchdog" safety monitor, user memory for up to 100 test routines, arc detection and available interface package with IEEE-488.2, RS232 and direct test results printout capability.

The intelligent choice for operator safety

At Vitrek, operator safety is not just a concern, it is a relentless commitment. So we engineered the 944i with a high-impact injection molded front-end with no exposed metal parts to present a potential user shock hazard. We put in an internal "watchdog" safety monitor which automatically shuts down the output in the event of an inadvertent disconnect. And the Vitrek 944i boasts an impressive user programmable "Arc Detection" circuit, capable of shutting down the high voltage system within microseconds.

For added peace of mind the onboard diagnostics of the 944i execute a complete check of all safety systems every time you power up.

World Class technology means reliability you can depend on

The Vitrek 944i provides increased performance and reliability when compared to discrete testers. Extensive use of ASIC, PAL and built-in test technologies make possible this high level of functionality with far fewer components than existing instruments.

If your product requires one of these labels. . .

Demand that the Vitrek 944i label is on your Dielectric Analyzer. Only Vitrek minimizes overall test time and maximizes system reliability by combining four dielectric test functions in one unit.

It would hardly seem possible that a slow dielectric tester could back-up an entire production line. Yet with today's imposing 100% dielectric inspection requirements, the antiquated manual hipot tester is often the manufacturing bottleneck. Imagine the consequences of the tester breaking down or worse yet passing a defective product.

Just a few good reasons why each 944i Multi-function Dielectric Analyzer is equipped with two high performance microprocessors - to speed testing throughput and to execute exhaustive onboard diagnostics. Simply put, the Vitrek 944i provides the fastest, most accurate and reliable test results available. And it does it for a surprisingly modest price.

Powerful yet easy to use

We've harnessed the power of the microprocessor to make this dynamic instrument simple and easy to operate. Specifically, the extensive use of menu prompting walks you through the test setup. Built-in user memory allows you to store, recall and edit up to 100 different test programs. All test parameters are fully user selectable - no hidden switches or pots to hamper operation.

The result is a clean, streamlined panel design that demands noticeably less effort to operate - and the unrivaled power to execute customized test routines at the touch of a button.

Capacity to excel

With its full 35mA sourcing capability, the Vitrek 944i possesses the horsepower required to drive today's capacitively filtered circuits - loads which the 10mA to 25mA units just can't handle. Yet, what sets the 944i apart from the rest is its ability to discriminate between in-phase and out-of-phase leakage currents.

This advanced feature enables you to measure the resistive current associated with insulation breakdown, the reactive component or the combination of both.

Vitrek Service & Support: Finest in the Instrument Industry

At Vitrek, we manufacture human engineered automated test instruments. But that's not where our commitment stops. We back you up with a worldwide network of Vitrek support centers. And we provide support by phone or email for any need or questions that may arise.

This all adds up to one point - you'll enjoy doing business with Vitrek. Our service technicians, application engineers and customer service satisfaction people just wouldn't have it any other way.

Features / Benefits

  • AC & DC Hipot plus Insulation Resistance and Ground Continuity in a single feature rich compact
    instrument
  • AC/DC Hipot to 5Vk standard, 7 and 10k VDC optional
  • Easy to use Full Menu Prompting, walks you through the test set-up
  • Test Database stores 200 test steps for easy recall
  • Powerful 35mA Sourcing capability
  • On-board Phase Detection measures "total", "real" or "imaginary" Leakage Current
  • Continuous Output Frequencies from 50 to 100Hz
  • IR measurement to 65 gigohms
  • Continuity measurement to 1 milliohm
  • Leakage Current measurement to 1 nanoamp
  • Optional IF-4 Interface package includes GPIB, RS232, VICL & Printer Ports
  • Separate Ramp & Dwell, Min & Max current limits
  • Covers-on Automatic Calibration
  • Built-in controls for HV Scanner and Ground Bond Tester
  • CD Mark Certified to EN61010
  • Includes NIST Traceable Certification
  • Three Year Extended Warranty
  • Made in the USA

Specifications

AC DIELECTRIC MODE

AC Voltage Output

 Range  Current   Accuracy
100 to 500VAC 20mA     0.7% of output ?7.5V
500 to 5000VAC  35mA 0.7% of output ?30V

Output Frequency:  50Hz to 100Hz with 1Hz resolution independent of line power frequency.

Output Waveform:  Digitally synthesized, low distortion sinewave.

AC Current Sensing (AC Leakage)

Range    Resolution   Accuracy
13.00uA 10nA 0.5% of reading ?1uA
1.300mA 1uA 0.5% of reading ?10uA
40.0mA 100uA 0.5% of reading ?500uA

Current Sensing:  Resistive current (in-phase), reactive current (out-of-phase) or total current.

Voltage Coefficient:  100pA/V    

DC DIELECTRIC MODE

DC Voltage Output

Range  Current   Accuracy
100 to 500VDC 20mA  0.3% of output ?6V
500 to 5,000VDC 35mA 0.3% of output +?5V
5,000 to 7,000VDC* 5mA 0.3% of output  ?5V

Requires option *DC-7

DC Voltage Output with option DC-10
500 to 10,000VDC       5mA        1% of output + 50V

DC Current Sensing (DC Leakage)

Range Resolution Accuracy
1.800uA    1nA  0.4% of reading ?6nA
18.00uA 10nA 0.3% of reading ?30nA
1.800mA 1uA 0.3% of reading ?30uA
40.0mA 10uA 0.3% of reading ?300uA

Voltage Coefficient:  15pA/V

HIGH RESISTANCE MODE
(Insulation Resistance)

Resistance Range Minimum Resistance Test Voltage
2.6GΩ 100KΩ   100VDC
13GΩ   500KΩ 500VDC
26GΩ 1MΩ  1,000VDC
65GΩ  2.5MΩ 2,500VDC

Accuracy: Combined uncertainty of DC Voltage Output + Current Measurement Accuracy

LOW RESISTANCE MODE
(Earth Continuity)

Full Scale 8.5Ω  Maximum including lead resistance.
Resolution 1mΩ    
Test Current 50mA
Lead Compensation Automatic offset circuit "nulls" out two wire lead resistance
Accuracy 0.3% of reading + 30mΩ (up to 5Ω)

General Specifications

Arc Detection Sensitivity User selectable 3 microseconds to 30 microseconds
Display 40 character alphanumeric LCD
High Voltage Indicator Lightning bolt illuminated by independent output voltage monitor.  Audible alarm sounds prior to voltage output.
Voltage Resolution 1 volt AC or DC
Dwell Time Accuracy +2 seconds/minute, -500mS/minute
Slew Rate 10 volts/second to 1999 volts/second, user selectable
Limits Test automatically aborts when digitally stored minimum or maximum limit is exceeded.
Remote Interface Option "IF-4"
GPIB - IEE-488.2 Subset compatible
RS232C - 1200, 2400, 4800, 9600 Baud rate selectable
Parallel - IBM/Centronics compatible
VICL - Vitrek auxiliary unit control loop
Accuracy
Stated specifications are bipolar and apply for one year within ?5?span style="font-size: 10.0pt; mso-bidi-font-size: 12.0pt">C from calibration temperature (20 to 25?span style="font-size: 10.0pt; mso-bidi-font-size: 12.0pt">C).
Temperature Coefficient
0.1 times stated accuracy per 蚓 beyond calibration temperature window
Calibration
Traceable to the National Institute of Standards and Technology (NIST). Automatic "Covers on" calibration, internal or external.  Calibration constants stored in non-volatile memory.
Warm up Time 30 minutes
Operating Temperature 0蚓 to 50蚓
Humidity 90% RH, 0?to 40蚓
Power 115/230 VAC ?/font> 10%, 350 VA Max
Dimensions 89mm H x 432mm W x 457mm D (3.5" H x 17" W x 18" D)
Weight 14Kg (34 lb.) Net / 18Kg (40 lb.) Shipping