4600A Series Precision High Voltage Meter
HV Standards Lab in a Box When it comes to making precise high voltage measurements, there is nothing faster or easier than the Vitrek 4600A Series.  With its unrivaled accuracy and direct reading capability, the 4600A Series sets the performance standard for precision high voltage measurement.  For quick and accurate calibration of Hipot testers (dielectric strength testers) and other high voltage sources, the 4600A Series can be beat.  Whether you are making direct or alternating high voltage measurements, the 4600A Series is uniquely equipped for the job.  For rock solid DC performance, the 4600A Series' multi-slope integrating analog to digital converter provides fast, stable readings even in the presence of noise and ripple.  For accurate AC measurement with either sine wave or distorted waveforms all of the 4600A Series units feature true RMS technology.  The 4600A Series is equipped with two voltage scales, so you can cover a broader span of voltages and select the ideal range for optimum performance.  

Superior Performance by Design 

Measurement integrity is our bottom line ?whether you are on a production line manufacturing high voltage medical imaging equipment, in a national laboratory conducting ground breaking research or in the cal lab certifying test equipment, it is essential to choose your tools carefully.  Take the Vitrek 4600A Series, for instance.  This rugged, compact instrument is equally at home in the lab or in the field.  Underneath its attractive, high impact, injection molded enclosure, lies a steel Faraday cage - to guard against the introduction of uncertainties due to stray EMFs, which are often present in high voltage applications.  The modular construction of the 4600A Series probe features a removable probe tip, which can be replaced with a variety of available HV leads to provide the ideal connection for your application.  The high impedance attenuator design of the 4600A Series balances offsetting power and voltage coefficients to deliver lab grade performance in a benchtop package.  The 4600A Series attenuator also features a patented non-inductive design to minimize field effects and provide enhanced AC performance.  For people on the go ?the 4600A Series is available with a standard carrying handle and an optional battery pack.  For systems applications the 4600A Series is available with rack mount adaptor and rear panel inputs.   The Vitrek 4600A Series is not only built tough, it is built to last ?and it is backed by our Extended Three Year Warranty.  Designed and manufactured right here in the USA, the 4600A Series is backed by the best customer service team in the northern hemisphere.  Isn't about time to put Vitrek legendary reputation for quality and reliability to work for your team?

Features / Benefits

  • Precise HV measurement up to 20kV, 40kV or 70kV
  • Ideal for Calibrating Hipot Testers, HV power supplies, Insulation Testers, HV Transformers and more
  • High Accuracy DC and True RMS AC capability
  • Compact, Rugged, Lightweight ?Benchtop/Portable Enclosure
  • Modular Probe Tip accepts a Variety of Test Lead Terminations
  • Multi-slope Integration provides Fast, Stable readings with outstanding noise rejection
  • Large, Bright, Easy to read 20,000 Count LED Display
  • Two HV Measurement Ranges provide optimum performance over a broad range of voltages
  • CE mark Certified to EN61010
  • Three Year Extended Warranty

 

Specifications

DC Voltage Measurements

Ranges

4600A -  2kV, 20kV

4640A -  2kV, 40kV

4670A -  20kV, 70kV

Resolution

100mV on 2KV, 1V on 20kV, 10V on 40kV and 70kV ranges

Accuracy

(valid 1 year from date of calibration at 23˚C ?C)
0.04% of reading ?.02% of range

Temperature coefficient 0.005% of reading per ˚C from 10 to 20 and 26 to 40
Voltage Coefficient 0.1ppm/V
Maximum Inputs

(HV Probe Tip)

0kVDC 4600A,

5kVDC 4640A,

0kVDC 4670A, (common terminal)

500VDC max

AC Voltage Measurements (True RMS)

Ranges 4600A - 2kV, 20kV
4640A - 2kV, 40kV
4670A - 20kV, 70kV
Resolution 100mV on 2kV, 1V on 20kV, 10V on 40kV and 70kV ranges
Accuracy

(1 year from date of calibration at 23˚C)
2kV Range:   20 to 100Hz 0.07% of reading 0.1% of range

                    100 to 400Hz 0.4% of reading 0.2% of range
20kV Range:  20 to 100Hz 0.2% of reading 0.1% of range
40kV Range:  50 to 60Hz 0.4% of reading 0.15% of range
70kV Range:  50 to 60Hz 0.4% of reading 0.15% of range

Temperature coefficient 0.01% of reading per ˚C from
10 to 20 蚓 and 26 to 40 蚓
Voltage Coefficient

0.2ppm/V on 2kV, 20kV and 40kV ranges 0.15ppm/V on 70kV range

Maximum Inputs

(at HV probe tip) 20kV peak AC (15kVRMS sine) 4600A,
40kV peak AC (25kVRMS sine) 4640A,

50kV peak AC (35kVRMS sine) 4670A,

(at common terminal) 500 Volts peak AC

General Specifications

Display Resolution 4 digits (19.999 count)
Input Impedance 200MΩ, 400MΩ with 4670A
Reading Rate 300 milli-seconds
NMRR 50db at 50 or 60 Hz
CMRR 50db
Operating Temperature 10 to 40
Humidity 85% RH
Safety CE mark certified to EN 61010
Power 115/230VAC 0%, 50/60Hz, 10VA Maximum
Dimensions 292mmW x 203mmD x 83mmH (11.5"W x 8"D x 3"H)
Probe Dimensions
Model Length Overall Creepage Distance
4600A 20kV Probe 406mm (16) 280mm (11)
4640A 40kV Probe 457mm (18) 380mm (15)
4670A kV70 Probe 508mm (20) 482mm (19)
Weight 2.5kg (5.5 lbs) Net, 5kg (11 lbs) Shipping
Standard Accessories High Voltage probe with removable probe tip, common input test lead, TL-46PK high voltage test lead variety kit, power cord, operating manual, NIST traceable - ANSI/NCSL Z540 certificate of calibration
Warranty 3 year extended warranty with no charge registration and annual factory calibration, one year warranty standard.
3 Probes